EasyManuals Logo
Home>Mindray>Diagnostic Equipment>TEX20

Mindray TEX20 Service Manual

Mindray TEX20
445 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #371 background imageLoading...
Page #371 background image
Revision:1.0(2023-01-12)
372
Involved FRU
Main Board Assembly and Probe Board assembly
Severity
Suspension
Handling Level
Contact the device provider for onsite handling.
Error Mechanism
PHV2N overcurrent or overpower protection
Possible Cause
1. The PHV voltage load causes the abnormal voltage output.
2. The power circuit of the PHV board fails.
Resolve Path
Cross-test the probe and the probe groove. If the probe and the probe board are normal, apply to
replace the main unit assembly.
Solution
1. Restart the machine, and check whether the fault is rectified.
2. Check whether the fault exists on a specific probe groove. If yes, clean the probe groove. If the
fault persists after cleaning, replace the probe board assembly. For details, refer to 7.9.2
Disassembly and Assembly.
3. Check whether the fault exists on a specific probe. If yes, replace the probe.
4. If the fault appears on all probes, cross-test or replace the main board. For details, refer to
7.13.2
Disassembly and Assembly.
88..22..55..1188 P1003: Transmit voltage error. The image cannot be displayed
correctly. Please try to reboot the device.
Failure Code
P1003
Fault Description
Diagnostic Ultrasound System
Service Manual
8 Troubleshooting

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals