88..33..77..77 The main display is normal. The external HDMI resolution is
inappropriate.
Fault Description
The main display is normal. The external HDMI resolution is inappropriate.
Involved FRU
/
Severity
Maintenance required
Handling Level
Handle the error yourself.
Possible Cause
The external display resolution fails to be adapted to.
Resolve Path
Manually set the output resolution of an external display in the Setup menu.
Solution
Enter the Setup screen. Choose System>Output. In the Resolution area of the Digital Output
Size menu, select different resolutions for testing.
88..33..77..88 The main display is normal and the HDMI external display has strips
or blinks
Fault Description
The main display is normal and the HDMI external display has strips or blinks
Involved FRU
/
Severity
Maintenance required