8-19
Arrh. event Range Default Step Unit
Vtac Rate 100 to 200
Adult, pediatric: 130
Neonate: 160
5 bpm
Vtac PVCs 3 to 99 6 1 /min
Pause Time 1.5, 2.0,2.5 2 / s
Vbrd PVCs 3 to 99 5 1 /min
Vbrd Rate 15 to 60 40 5 bpm
8.7.4 Setting the Extended Arrhythmia
The following arrhythmia events are defined as extended arrhythmia:
Extreme Tachy
Extreme Brady
Vent. Brady
Nonsus. Vtac
Multif. PVC
Irr. Rhythm
Pause
Afib
You can select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required passwordselect [Alarm Setup >>], and set [Extended Arrh.] to [Enable] or
[Disable]. When [Extended Arrh.] is set to [Disable], the patient monitor does not analysis
the extended arrhythmia events and corresponding alarms are not given.
CAUTION
z Set [Extended Arrh.] to [Disable] when the patient monitor is connected to the
Central Monitoring System of version prior to 06.01.00. Failure to do so may cause
the Central Monitoring System unable to display extended arrhythmia related
alarms normally when extended arrhythmia occurs.
8.7.5 Reviewing Arrhythmia Events
Please refer to 18 Review.