7.5.3 Setting Alarm Delay Time 
You can set the alarm delay time for over-limit alarms of continuously measured parameters. 
If the alarm-triggered condition disappears within the delay time, the patient monitor will 
not give the alarm. You can set the [Alarm Delay] in the [Others] window of [Alarm
Setup] menu.   
 
To set the alarm delay time,   
1. Select [Main Menu] [Maintenance >>] [User Maintenance >>]. Enter the 
required password and then select [Ok]. 
2. Select [Alarm Setup >>][Alarm Delay]. 
 
Alarm delay is not applied to the following physiological alarms:   
  Apnea 
  ST alarms 
  Arrhythmia alarms 
  ECG Weak Signal 
  Resp Artifact 
  No Pulse 
  Measurements of noncontinuous parameters over alarm limits 
  HR over alarm limits 
 
You can set [Apnea Delay] and [ST Alarm Delay] separately. 
 
7.5.4 Setting SpO
2
 Technical Alarm Delay 
You can set [Tech. Alarm Delay] in the [Others] tab of the [Alarm Setup] menu. The 
options are [Off], [5s], [10s] and [15s]. The delay is effective to the following technical 
alarms: SpO
2
 Sensor Off, SpO
2
 Too Much Light, SpO
2
 Low Signal and SpO
2
 Interference.