9-2 
9.3 Understanding the Resp Display 
 
 
 
 
 
(1) Gain 
(2) Resp lead label 
(3) Respiration rate 
(4) RR source 
 
By selecting the waveform area or parameter area, you can enter the [Resp Waveform] menu. 
By selecting the Resp parameter window, you can enter the [Resp Setup] menu. 
 
NOTE
z  Respiration monitoring is not for use on the patients who are very active, as this 
will cause false alarms. 
 
 
9.4 Placing Resp Electrodes 
As the skin is a poor conductor of electricity, preparing the skin is necessary for a good 
Respiration signal. You can refer to 8 Monitoring ECG for how to prepare the skin. 
 
As the Respiration measurement adopts the standard ECG electrode placement, you can use 
different ECG cables (3-lead or 5-lead). Since the respiration signal is measured between two 
ECG electrodes, if a standard ECG electrode placement is applied, the two electrodes should 
be RA and LA of ECG Lead I, or RA and LL of ECG Lead II. 
 
NOTE
z  To optimize the respiration waveform, place the RA and LA electrodes 
horizontally when monitoring respiration with ECG Lead I; place the RA and LL 
electrodes diagonally when monitoring respiration with ECG Lead II. 
 
 
(4)
(1) 
(2) 
(3)