EasyManuals Logo

Mindray uMEC15 User Manual

Mindray uMEC15
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
7-14
You can also temporarily prolong the alarm pause time after the monitor entersthe alarm
paused status:
1. Select [Main Menu]Ä—[Maintenance >>]Ä—[User Maintenance >>]Ä—enter the
required passwordÄ—[Alarm Setup >>].
2. In the [Alarm Setup] menu, set the [Max. Alarm Pause 15min] to [Enable].
3. In the physiological alarm area, select a proper time in the [Alarm Pause Time] menu.
7.7 Switching Off All Alarms
If [Alarm Pause Time] is set to [Permanent], the patient monitor enters into the alarm off
status after the
hardkey is pressed. During the alarm off status,
 As for physiological alarms, no alarm lamps flash and no alarms are sounded.
 As for physiological alarms, no numeric and alarm limit flash.
 No physiological alarm messages are shown.
 [Alarm Off] is displayed in the physiological alarm area with red background.
 As for technical alarms, no alarms are sounded.
 The
alarm off symbol is displayed in the alarm symbol area.
You can cancel the alarm off status by pressing the
hardkey.
WARNING
z Pausing or switching off alarms may result in a hazard to the patient. Please be
very careful.
NOTE
z [Max. Alarm Pause 15min] is configured to [Disable] by default. In this case, you
cannot prolong the pause time.The prolonged pause time is only effective to the
current paused alarms.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC15 and is the answer not in the manual?

Mindray uMEC15 Specifications

General IconGeneral
BrandMindray
ModeluMEC15
CategoryMedical Equipment
LanguageEnglish

Related product manuals