14-3 
 
14.4 Setting Up the Pressure Measurement 
1.  Plug the pressure cable into the IBP connector. 
2.  Prepare the flush solution. 
3.  Flush the system to exhaust all air from the tubing. Ensure that the transducer and 
stopcocks are free of air bubbles. 
 
 WARNING 
z  If air bubbles appear in the tubing system, flush the system with the infusion 
solution again. Air bubble may lead to wrong pressure reading. 
 
4.  Connect the pressure line to the patient catheter. 
5.  Position the transducer so that it is level with the heart, approximately at the level of the 
midaxillary line. 
6.  Select the appropriate label. 
7.  Zero the transducer. After a successful zeroing, turn off the stopcock to the atmosphere 
and turn on the stopcock to the patient. 
 
 
 
 
Heparinized 
fluid bag 
 
Flushing set 
Valve 
Pressure 
transduce
 
Sto
cock
Adapter cable for using 
dis
osable transducers 
Pressure line
Monito