7-18
7.11.3 Resetting Care Group Alarms
You can reset the alarms presented on the viewed bed by pressing the [Alarm Reset] from
the current monitor’s [View Other Patient] window. To enable this function:
1. Select [Main Menu]Ä—[Maintenance>>]Ä—[User Maintenance>>]Ä—enter the
required passwordÄ—[Alarm Setup>>].
2. In the [Alarm Setup] menu, set [Reset Other Bed’s Alarms] to [On].
The alarms presented on the current monitor can also be reset from another monitor viewing
this monitor. To do so, proceed as follows:
1. In the current monitor, select [Main Menu]Ä—[Maintenance>>]Ä—[User
Maintenance>>]Ä—enter the required passwordÄ—[Alarm Setup>>].
2. In the [Alarm Setup] menu, set [Alarm Reset By Other Bed] to [On].
3. In the other monitor, select the [Alarm Reset] button from the [View Other Patient]
window.
WARNING
z Resetting care group alarms may cause a potential hazard. Please act with caution.
7.12 When an Alarm Occurs
When an alarm occurs, observe the following steps and take proper actions:
1. Check the patient’s condition.
2. Confirm the alarming parameter or alarm category.
3. Identify the source of the alarm.
4. Take proper action to eliminate the alarm condition.
5. Make sure the alarm condition is corrected.
For troubleshooting specific alarms, see appendix D Alarm Messages.