EasyManuals Logo

Mindray uMEC15 User Manual

Mindray uMEC15
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #204 background imageLoading...
Page #204 background image
24-2
NOTE
z To clean or disinfect reusable accessories, refer to the instructions delivered with
the accessories.
z Avoid the external connectors and thermovent during cleaning or disinfection
procedures.
24.2 Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of
dust and sand in your place, the equipment should be cleaned more frequently. Before
cleaning the equipment, consult your hospital’s regulations for cleaning the equipment.
Recommended cleaning agents are:
Product Name Product Type Ingredient
Water Liquid /
Ethanol Liquid Ethanol 70%
Isopropyl alcohol Liquid Isopropyl alcohol 70%
Sodium hypochlorite bleach Liquid Sodium hypochlorite bleach 10%
Hydrogen peroxide Liquid Hydrogen peroxide 3%
1-Propanol Liquid 1-Propanol 50%
Rely+On
TM
Virkon
®
Powder
Used as 1% solution (prepared the
concentration directed by the
accompanying instructions for use)
Biocidal active:
 Pentapotassium bis
(peroxymonosulphate) bis (sulphate)
(500g/kg)
 Contains dipotassium
peroxodisulphate
Super Sani-cloth
®
Germicidal Disposable Wipe
Wipe
Quaternary ammonium chloride 0.5%
Isopropyl alcohol 55%

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC15 and is the answer not in the manual?

Mindray uMEC15 Specifications

General IconGeneral
BrandMindray
ModeluMEC15
CategoryMedical Equipment
LanguageEnglish

Related product manuals