16-1 
16 Monitoring Carbon Dioxide 
(for uMEC12/uMEC15) 
 
16.1 Introduction 
CO
2 
monitoring is a continuous, non-invasive technique for determining the concentration of 
CO
2
 in the patient’ airway by measuring the absorption of infrared (IR) light of specific 
wavelengths. CO
2 
has its own absorption characteristic and the amount of light passing the 
gas probe depends on the concentration of the measured CO
2. 
When a specific band of IR 
light is passed through respiratory gas samples, some of IR light will be absorbed by the CO
2
 
molecules. The amount of IR light transmitted after it has been passed through the respiratory 
gas sample is measured with a photodetector. From the amount of IR light measured, the 
concentration of CO
2
 is calculated. 
 
CO
2 measurements are used to monitor the patient’s respiratory status. 
 
CO
2 monitoring is intended for adult, pediatric and neonatal patients. 
 
16.2 Safety 
 
WARNING
z  Remove the airway sampling line from the patient’s airway while nebulized 
medications are being delivered. 
z  Leakage in the breathing or sampling system may cause the displayed EtCO
values to be significantly low. Always make sure that all components are securely 
connected. 
z  EtCO
values measured from the CO
module may differ from those of from the 
blood gas analysis. 
z  Route all tubing away from the patient’s throat to avoid strangulation. 
z  Inspect the airway adapter for a tight connection and proper operation before 
attaching it to the patient. 
z  Squeezing or bending the sampling line during the CO
measurement may cause 
inaccurate CO
reading or no reading. 
z  Avoid mechanical shock to the CO
module configuring the paramagnetic oxygen 
sensor.