8-20 
 
8.8 About QT/QTc Interval Monitoring   
The QT interval is defined as the time between the beginning of the Q-wave and the end of 
the T-wave. It measures the total duration of the depolarization (QRS duration) and 
repolarization (ST-T) phases of the ventricular action potential. QT interval monitoring can 
assist in the detection of prolonged QT interval syndrome. 
The QT interval has an inverse relationship to heart rate. Faster heart rates shorten the QT 
interval and slower heart rates prolong the QT interval. Therefore, several formulas can be 
used to correct the QT interval for heart rate. The heart rate corrected QT interval is 
abbreviated as QTc.   
QT/QTc interval monitoring is intended for adult, pediatric, and neonatal patients. 
 
 
8.8.1 QT/QTc Monitoring Limitations 
Some conditions may make it difficult to achieve reliable QT/QTc monitoring, for example:   
  R-wave amplitudes are too low 
  Too many ventricular beats 
  RR intervals are not stable 
  A high heart rate causes the P-wave to encroach on the end of the previous T-wave 
  The T-wave is very flat or T-wave are not well defined 
  The end of the T-wave is difficult to define because of the presence of U-waves 
  QTc measurements are not stable 
  In the presence of noise, asystole, ventricular fibrillation, and ECG lead off 
 
For these cases you should select a lead with good T-wave amplitude and no visible flutter 
activity, and without a predominant U-wave or P-wave. 
 
Some conditions such as left or right bundle branch block or hypertrophy can lead to a 
widened QRS complex. If a long QTc is observed you should verify it to ensure that it is not 
caused by QRS widening. 
 
Because normal beats followed by ventricular beats are not included in the analysis, no QT 
measurement will be generated in the presence of a bigeminy rhythm. 
 
If the heart rate is extremely high (over 150bpm for adults and over 180bpm for pediatrics 
and neonates), QT will not be measured. When the heart rate changes, it can take several 
minutes for the QT interval to stabilize. For reliable QTc calculation it is important to avoid 
the region where the heart rate is changing.