7-7 
 
7.4.5 Setting the Reminder Tones 
When the alarm volume is set to zero, or the alarm tone is reset or switched off, the patient 
monitor issues a periodical reminder tone. To set the reminder tones:   
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the 
required password. 
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu. 
3. Set the [Reminder Tones] to [On], [Off] or [Re-alarm]. When [Re-alarm] is selected, 
the current physiological alarms and technical alarms marked with “Ĝ” will be 
re-generated after the [Reminder Interval] if the alarm condition persists. 
 
To set the interval between reminder tones, select [Reminder Interval] and toggle between 
[1min], [2min] and [3min]. 
 
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm 
reminder tones, select [Main Menu][Alarm Setup >>][Others] or the [Alarm Setup] 
QuickKey[Others]. Then, select [Reminder Vol] and toggle between [High], [Medium] 
and [Low]. 
 
 
7.5 Understanding the Alarm Setup Menu 
Select the [Alarm Setup] QuickKey or [Main Menu][Alarm Setup >>] to enter the 
[Alarm Setup], where you can:   
  Set alarm properties for all parameters. 
  Change ST alarm settings. 
  Change arrhythmia alarm settings. 
  Set the threshold for some arrhythmia alarms. 
  Change other settings.