A-1
A Product Specifications
A.1 Monitor Safety Specifications
A.1.1 Classifications
The patient monitor is classified, according to IEC60601-1:
Type of protection against
electrical shock
Class I, equipment energized from an external and
internal electrical power source.
Degree of protection against
electrical shock
Type BF defibrillation proof for CO
2
monitoring.
Type CF defibrillation proof for ECG, RESP, TEMP,
SpO
2
, NIBP, IBP and C.O.
Mode of operation
Continuous
Degree of protection against
harmful ingress of water
IPX1 (Protected against vertically falling water drops)
A.1.2 Environmental Specifications
WARNING
z The equipment may not meet the performance specifications if stored or used
outside the specified temperature and humidity ranges.
Main unit
Item Operating conditions Storage conditions
Temperature (ć) 0 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (mmHg) 427.5 to 805.5 120 to 805.5
CO
2
module
Item Operating conditions Storage conditions
Temperature (ć) 5 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (mmHg) 430 to 790 430 to 790