A-13 
 
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure 
measurements measured with this device are in compliance with the Standard for 
Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and stardard 
deviation by comparing with intra-arterial or auscultatory measurements (depending on the 
configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff 
sound was used to determine the diastolic pressure. 
In neonatal mode, the blood pressure measurements measured with this device are in 
compliance with the American National Standard for Non-invasive sphygmomanometers 
(ISO 81060-2) in terms of mean error and stardard deviation by comparing with 
intra-arterial measurements (depending on the configuration) in a typical patient population. 
 
 
A.7.6 Temp
Standards Meet standard of ISO 80601-2-56 
Technique Thermal resistance 
Operating mode  Direct mode 
Measurement range  0 to 50 ć (32 to 122 ) 
Resolution 0.1 ć 
Accuracy ±0.1 ć(without probe) 
Refreshing rate  1 s 
Minimum time for accurate 
measurement 
Body surface: <100 s 
Body cavity: <80 s 
Transient response time 
Body surface probe : <45 s 
Body cavity probe: <45 s 
 
Alarm limit  Range  Step 
T1/T2 High  (low limit +1) to 50 ć 
0.1 ć 
T1/T2 Low  0.1 to (high limit -1) ć 
TD High  0.1 to 50 ć