D-2 
Measurement  Alarm messages  L Cause and solution 
Extreme Brady  H 
R on T  M* 
Run PVCs  L* 
PVCs/min  M* 
Multif.PVC  M* 
Bigeminy  M* 
Trigeminy  M* 
Tachy  M* 
Brady  M* 
Vent. Rhythm  M* 
Multif. PVC  M* 
Nonsus. Vtac  M* 
Pause  L* 
Resp 
Resp Apnea  H 
The respiration signal was so weak that the 
monitor cannot perform respiration analysis. 
Check the patient’s condition and the Resp 
connections. 
Resp Artifact  H 
The patient’s heartbeat has interfered with 
his respiration. Check the patient’s condition 
and the Resp connections. 
SpO
2
 
SpO
2
 Desat  H 
The SpO2 value has fallen below the 
desaturation alarm limit. Check the patient’s 
condition and check if the alarm limit 
settings are correct. 
No Pulse  H 
The pulse signal was so weak that the 
monitor cannot perform pulse analysis. 
Check the patient’s condition, SpO
2 
sensor 
and measurement site. 
CO
2
  CO
2
 Apnea  H 
The patient stops breathing, or the 
respiration signal was so weak that the 
monitor cannot perform respiration analysis. 
Check the patient’s condition and the RM 
connections.