EasyManuals Logo

Mindray uMEC15 User Manual

Mindray uMEC15
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #274 background imageLoading...
Page #274 background image
D-8
Measurement Alarm message L I Cause and solution
Recorder Vlt Low L C
supply. Restart the monitor.
Recorder Head Hot L C The recorder has been working for too long
time. Stop the recording and resume the
recording till the recorder’s printhead cools
down.
Rec Paper Wrong Pos. L A Re-load the recorder paper.
System System Watchdog Err H C An error occurred to the system. Restart the
monitor.
System Software Err H C
System CMOS Full H C
System CMOS Err H C
System FPGA Err H C
System Err N H C
N is within 2 to 12.
Other Bed
Disconnected
L A Check network connection.
PWR interrupted.
Check meas. state
L A Power supply failed accidently. Check the
measurements when the monitor restarts.
No CMS L A The monitor is disconnected from the CMS.
Check network connection.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC15 and is the answer not in the manual?

Mindray uMEC15 Specifications

General IconGeneral
BrandMindray
ModeluMEC15
CategoryMedical Equipment
LanguageEnglish

Related product manuals