7-10
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits
range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
NIBP-M
(Mean × 0.68
+ 8) mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean × 0.86
+ 35) mmHg
(Mean + 15)
or 95 mmHg
(whichever
is smaller)
Adult: 30 to 230
Pediatric: 30 to 165
Neonate: 25 to 105
Temp
T1 (T1 – 0.5)ć
(T1 –
0.5) ć
(T1 + 0.5)ć (T1 + 0.5)ć 1 to 49 ć
T2 (T2 – 0.5)ć
(T2 –
0.5) ć
(T2 + 0.5)ć (T2 + 0.5)ć 1 to 49 ć
TD
Same as the
default alarm
limit
Same as the
default
alarm limit
Same as the
default alarm
limit
Same as the
default
alarm limit
Same as the
measurement range
IBP:
RT
o
AP/
BAP
FAP/
LV/
P1-P4
(Arterial
ressure)
IBP-S
(SYS × 0.68
ˇ10) mmHg
(SYS – 15)
or 45mmHg
(whichever
is greater)
(SYS × 0.86
ˇ38) mmHg
(SYS + 15)
or
105mmHg
(whichever
is smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
IBP-D
(Dia × 0.68 +
6)mmHg
(Dia – 15) or
20mmHg
(whichever
is greater)
(Dia × 0.86 +
32)mmHg
(Dia + 15) or
80mmHg
(whichever
is smaller)
Adult: 25 to 225
Pediatric: 25 to 150
Neonate: 20 to 90
IBP-M
(Mean × 0.68
+ 8)mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean × 0.86
+ 35)mmHg
(Mean + 15)
or 95mmHg
(whichever
is smaller)
Adult: 30 to 245
Pediatric: 30 to 180
Neonate: 25 to 105
IBP: PA
IBP-S SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25
3 to 120mmHg IBP-D Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1-P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40mmHg