No. 99MBB091A
5 - 22
Operating procedure if a profile other than assessed profile MOTIF is specified
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
Length set
Length
4 5 6
7BS
AC
8 9
1 2
0
3
ESC
mm
0.3
.
(0.8 7.2mm)
F1 F2 F3 F4
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N
53
L 0.3mm
1
1
<Home screen>
[Condition]
2
<Measurement Condition Setup screen>
[N]
[L] (arbitrary length)
3
<Number (N) Setup window>
4
<Arbitrary Length Input screen>
Arbitrary length input filed
[Enter]
Touch the [Condition] button on the Home screen.
The Measurement Condition Setup screen will appear.
Touch the [N] button.
The Number (N) Setup window will appear.
Touch the [L] button on the Number (N) Setup window.
The Arbitrary Length Setup screen will appear.
Enter an arbitrary length in the filed, then touch the [Enter]
button.
An arbitrary length can be specified between 0.3 mm (.01")
and 12.5 mm (.49") with 0.1 mm (.01") increments. Specifica-
tion outside this range is invalid.
TIP For numerical data specification in input filed,
refer to section 2.2, “Outline of the Touch Panel
Inputting numeric values”.
NOTE Set an arbitrary evaluation length after deter-
mining a cutoff value and a filter setting.
The enterable range of evaluation lengths is
changed if the cutoff value or the filter is
changed.
The Measurement Condition Setup screen will be returned
to.