5. MODIFYING THE MEASUREMENT CONDITIONS
<Home screen>
Condition
JIS1994
GAUSS
L 0.3mm
(AUTO)
RANGE
Ra
Ry
CAL.
* * * *
* * * *
2000-01-01
m
m
CARD
F1 F2 F3 F4
N 0.3mm
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
0.3mm
0.25mm
2.5m
Page1/3
5
<Measurement Condition Setup screen>
[Home]
Touch the title bar in the Number (N) Setup window to accept
the selected number of sampling lengths.
The Number (N) Setup window closes.
Touch the [Home] button.
The Home screen will be returned to.
■ Operating procedure if assessed profile
(MOTIF.R/MOTIF.W) is specified
Length set
Length
4 5 6
7BS
AC
8 9
1 2
0
3
ESC
mm
3.3
.
(3.3 16mm)
s
B
EVA-L
Condition menu
Page1/3
STAND.
L
RE-CAL.
JIS'94
MOTIF.R
A
0.5mm
16.0mm 16.0mm
8m
PROFILE
FILTER
1 <Measurement Condition Setup screen>
[L] (Arbitrary length)
2
<Arbitrary Length Input screen>
Arbitrary length input field
Range that can be set
Touch [L] (Arbitrary length) on the Measurement Condition
Setup screen.
The Arbitrary Length Input screen is displayed.
Input an arbitrary length in the arbitrary length input field,
then touch the [Enter] button.
Even if a numeric value outside the range that can be set is
inputted, the input content will not be loaded.
TIP For imformation about the numeric value input
procedure, refer to “2.2 Outline of the Touch
Panel ■ Inputting numeric values”.
The Measurement Condition Setup screen will be restored.
NOTE An evaluation length is stored with respect to
the upper limit length (A). Therefore, if the
upper limit length (A) is changed, the system
will set the evaluation length that has been
previously stored at the changed upper limit
length (A).