5 - 43
No. 99MBB091A
5. MODIFYING THE MEASUREMENT CONDITIONS
■ Setting up mr calculation condition
F1 F2 F3 F4
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.83
7.12
2000-01-01
m
m
CARD
Rmr[c]
BASE
S.LEVEL
PEAK
10 %
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
CAL.
1.54
2000-01-01
%
m
CARD
Rmr[c]
Ra
(
P,10%
)
15
2
<Home screen>
3
<mr[c] Measurement Condition Setup window>
4
< (mr[c] Measurement Condition) Slice Level
Input screen>
Slice Level button
PEAK/BASE button
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
1
<Home screen>
[Parameter scroll]
F1 F2 F3 F4
F1 F2 F3 F4
Rmr[c] Slice Level
(-100 100%)
4 5 6
7
10
BS
AC
8 9
1 2
0 .
3
ESC
+/-
%
Parameter button
Touch the [Parameter scroll] button in the Home screen one
or more times until the parameter (mr[c]) appears in the
button, for which the calculation condition is to be set.
Touch the displayed parameter button.
The mr Measurement Condition Setup window will
appear.
If using the peak reference, touch the [PEAK] button, then
touch the [S.LEVEL] button. (To use the base reference,
touch the [BASE] button, then touch the [S.LEVEL] button.)
The mr Slice Level Input screen will appear.
To change the unit of cutoff levels, touch the [%] key. Each
time the key is touched, the unit is switched to either [%] or
[mm]. This section is described assuming that [%] is specified
as the unit.