5 - 54
No. 99MBB091A
Setting up HSC calculation condition
F1 F2 F3 F4
Rhsc Count Level
4 5 6
7
0.1
BS
AC
8 9
m
1 2
0 .
3
ESC
(0.0 999.9m)
1
2
3
4
[m]
<Home screen>
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
Condition
JIS1994
GAUSS
c
0.8mm
5
(AUTO)
RANGE
[Parameter scroll]
CAL.
2000-01-01
CARD
Condition
JIS1994
GAUSS
c
0.8mm
5
(AUTO)
RANGE
Ra
Rhsc
(
P,0.1m
)
****
1.54
/cm
m
Parameter button
<Home screen>
<HSC Measurement Condition Setup window>
Ra
Ry
CAL.
1.83
7.1
2000-01-01
m
m
CARD
Condition
JIS1994
GAUSS
c
0.8mm
5
(AUTO)
RANGE
Rhsc
BASE
C.LEVEL
PEAK
0.1
m
PEAK/BASE button
Count level button
<Count Level Input screen>
F1 F2 F3 F4
F1 F2 F3 F4
Touch the [Parameter scroll] button in the Home screen one
or more times until the parameter (HSC) appears in the
button, for which the calculation condition is to be set.
Touch the displayed parameter button.
The HSC Measurement Condition Setup window will
appear.
If using the peak reference, touch the [PEAK] button, then
touch the count level button. (To use the base reference, touch
the [BASE] button, then touch count level button.)
The Count Level Input screen will appear.
If the unit of count level is to be changed, touch the [µm]
([µin]) button. Each time this button is touched, the unit is
toggled between [µm] ([µin]) and [%].
Here, the procedure is explained using [µm] ([µin]) for the
unit of count level.