- 34 -
Calibrating the Sensor
(1)
Check the black mark sensor
position
Confirm the center of the black
mark on the media and the black
mark sensor position on the device
(number on the
indicator).
If the position of the black mark
and the black mark sensor position
do not align, rotate the Sensor
Position Adjustment Dial to adjust
the position (go to step 4).
The black mark sensor position is
set, at the factory, to 58.5 mm from
the right edge of the label liner.
Adjust the position of the “
” so it
is in the center of the black mark
of the media you are using.
Black mark
(on back side or paper)
Center of black mark
(2)
Check the gap sensor position
Confirm the center of the label
on the media and the gap sensor
position on the device (number on
the | indicator).
If the position of the label and
the gap sensor position do not
align, rotate the Sensor Position
Adjustment Dial to adjust the
position (go to step 4).
The gap sensor position is set, at
the factory, to 65 mm from the
right edge of the label liner.
Adjust the position of the “ | ” so it
is in the center of the labels of the
media you are using.
4
Rotate the Sensor Position Adjustment
Dial to adjust the sensor position.
A
B
D
C
Memo
z
If you rotate the Sensor Position Adjustment Dial in
direction A
→
the sensor position moves in the
direction C
direction B
→
the sensor position moves in the
direction D