Software Reference 118
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Run Info Tab (Continued)
Chip Expiration Date - The date when the chip expires.
Chip Remaining Samples - The number of samples that the chip
can process before reaching the maximum number of samples.
Chip Last Prime Date/Time - The date of the last prime for the
chip.
Chip Prime Count - The number of times the chip has been
primed.
The following information displayed on the Run Info Tab apply only
to Protein Clear HR assays:
Calibration Iterations - The number of attempts made to complete
the last phase of the calibration (i.e., Final Percent Purity
Obtained).
Calibration Ladder EP Mobility - The mobility of the 120 kDa
ladder peak during calibration.
Calibration Final DSR Applied - The Final DSR applied during
instrument calibration.
Final Percent Purity Obtained - The ratio of the corrected areas of
the 180 kDa and 200 kDa peaks in the calibration standard sample.
Calibration Max DSR Tested - The maximum DSR tested during
calibration.
Calibrated I0 - The current applied to electrode 0 on the chip during
the separation portion of the sample processing.
Calibrated I2 - The current applied to electrode 2 on the chip during
the separation portion of the sample processing.
Calibrated I6 - The current applied to electrode 6 on the chip during
the separation portion of the sample processing.
Calibrated V7 - The current applied to electrode 7 on the chip
during the separation portion of the sample processing.