Remote control commands
R&S
®
LCX Series
202User Manual 1179.2260.02 ─ 02
:LOG:INTerval 0
LOG:COUNt..................................................................................................................202
LOG:DURation.............................................................................................................. 202
LOG:FNAMe..................................................................................................................202
LOG:INTerval.................................................................................................................203
LOG:MODE...................................................................................................................203
LOG:STIMe...................................................................................................................204
LOG[:STATe]................................................................................................................. 204
LOG:COUNt <Sample Count>[, <Return min or max>]
LOG:COUNt? [<Return min or max>]
Sets the number of measurement readings in count mode. To set the mode, use the
command LOG:MODE on page 203 to be captured.
Setting parameters:
<Sample Count> Range: 1 to 10E+6
*RST: 10
Parameters for setting and query:
<Return min or max> MIN | MINimum | MAX | MAXimum
Example: See Example "Logging data settings" on page 201.
Manual operation: See "Count" on page 97
LOG:DURation <Logging Duration>[, <Return min or max>]
LOG:DURation? [<Return min or max>]
Defines the duration of logging for the measurement in span and duration mode. To set
the mode, use the command LOG:MODE on page 203.
Setting parameters:
<Logging Duration> Numeric value in seconds.
Range: 0 to 349199
*RST: 60
Parameters for setting and query:
<Return min or max> MIN | MINimum | MAX | MAXimum
Example: See Example "Logging data settings" on page 201.
Manual operation: See "Duration" on page 97
LOG:FNAMe <Logging File Name>
Sets the file name and path for the storing the data recorded during data logging.
LOG subsystem

Table of Contents

Other manuals for R&S LCX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S LCX Series and is the answer not in the manual?

R&S LCX Series Specifications

General IconGeneral
BrandR&S
ModelLCX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals