2. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown
in
Figure 4.7
.
3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal
Output Terminal" of the probe, as shown in
Figure 4.7
.
Figure 4.7 Using the Compensation Signal
4. Set the probe ratio based on the attenuation of the probe, and then click or tap
> Auto.
5. Observe the waveform on the display. In normal condition, you should see a
square waveform similar to the waveform shown in the figure below.
Figure 4.8 Square Waveform Signal
6. Test the other channels in the same way. If you see the waveform, but the square
wave is not shaped correctly as shown above, perform the procedure described in
Probe Compensation
. If you do not see the waveform, perform the inspection
again.
WARNING
Quick Start
HDO1000 User Guide
16
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.