RIGOL 2 Command System
2-86 MSO2000A/DS2000A Programming Guide
:CURSor:TRACk:XDELta?
Syntax
:CURSor:TRACk:XDELta?
Description
Query the difference X between the X values at cursor A and cursor B in track cursor measurement and
the unit is s.
Explanation
For the horizontal position of cursor A, refer to the
:CURSor:TRACk:CAX command.
For the horizontal position of cursor B, refer to the
:CURSor:TRACk:CBX command.
Return Format
The query returns the current difference X in scientific notation.
:CURSor:TRACk:YDELta?
Syntax
:CURSor:TRACk:YDELta?
Description
Query the difference Y between the Y values at cursor A and cursor B in track cursor measurement.
Explanation
For the vertical position of cursor A, refer to the
:CURSor:TRACk:CAY? command.
For the vertical position of cursor B, refer to the
:CURSor:TRACk:CBY? command.
The unit is determined by the unit (refer to the
:CHANnel<n>:UNITs command) of the current signal
source.
Return Format
The query returns the current difference Y in scientific notation.