Common Measurement Settings
R&S
®
FSW
370User Manual 1173.9411.02 ─ 19
"Table"
Defines the conversion loss via the table selected from the list. Pre-
defined conversion loss tables are often provided with the external
mixer and can be imported to the R&S FSW. Alternatively, you can
define your own conversion loss tables. Imported tables are checked
for compatibility with the current settings before being assigned.
Conversion loss tables are configured and managed in the Managing
Conversion Loss Tables tab.
For details on conversion loss tables, see "Conversion Loss Tables"
on page 361.
For details on importing tables, see "Import Table" on page 373.
Remote command:
Average for range 1:
[SENSe:]MIXer:LOSS[:LOW] on page 874
Table for range 1:
[SENSe:]MIXer:LOSS:TABLe[:LOW] on page 873
Average for range 2:
[SENSe:]MIXer:LOSS:HIGH on page 873
Table for range 2:
[SENSe:]MIXer:LOSS:TABLe:HIGH on page 873
Basic Settings
The basic settings concern general use of an external mixer. They are only available if
the External Mixer State is "On".
LO Level......................................................................................................................370
Signal ID......................................................................................................................371
Auto ID........................................................................................................................ 371
Auto ID Threshold....................................................................................................... 371
Bias Settings............................................................................................................... 371
└ Write to <CVL table name>...........................................................................372
LO Level
Defines the LO level of the external mixer's LO port. Possible values are from 13.0
dBm to 17.0 dBm in 0.1 dB steps. Default value is 15.5 dB.
Remote command:
[SENSe:]MIXer:LOPower on page 869