Remote Commands
R&S
®
FSW
868User Manual 1173.9411.02 ─ 19
Basic Settings....................................................................................................... 868
Mixer Settings....................................................................................................... 870
Conversion Loss Table Settings........................................................................... 874
Programming Example: Working with an External Mixer......................................878
Basic Settings
The basic settings concern general usage of an external mixer.
[SENSe:]MIXer[:STATe]..................................................................................................868
[SENSe:]MIXer:BIAS:HIGH.............................................................................................868
[SENSe:]MIXer:BIAS[:LOW]............................................................................................868
[SENSe:]MIXer:LOPower................................................................................................869
[SENSe:]MIXer:SIGNal...................................................................................................869
[SENSe:]MIXer:THReshold............................................................................................. 869
[SENSe:]MIXer[:STATe] <State>
Activates or deactivates the use of a connected external mixer as input for the mea-
surement. This command is only available if the R&S FSW-B21 option is installed and
an external mixer is connected.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MIX ON
Manual operation: See "External Mixer State" on page 367
[SENSe:]MIXer:BIAS:HIGH <BiasSetting>
This command defines the bias current for the high (second) range.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 868).
Parameters:
<BiasSetting> *RST: 0.0 A
Default unit: A
Manual operation: See "Bias Settings" on page 371
[SENSe:]MIXer:BIAS[:LOW] <BiasSetting>
This command defines the bias current for the low (first) range.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 868).
Parameters:
<BiasSetting> *RST: 0.0 A
Default unit: A
Setting Basic Measurement Parameters