Remote Commands
R&S
®
FSW
870User Manual 1173.9411.02 ─ 19
Manual operation: See "Auto ID Threshold" on page 371
Mixer Settings
The following commands are required to configure the band and specific mixer set-
tings.
[SENSe:]MIXer:FREQuency:HANDover........................................................................... 870
[SENSe:]MIXer:FREQuency:STARt?................................................................................870
[SENSe:]MIXer:FREQuency:STOP?................................................................................ 871
[SENSe:]MIXer:HARMonic:BAND:PRESet........................................................................871
[SENSe:]MIXer:HARMonic:BAND[:VALue]........................................................................871
[SENSe:]MIXer:HARMonic:HIGH:STATe..........................................................................872
[SENSe:]MIXer:HARMonic:HIGH[:VALue]........................................................................ 872
[SENSe:]MIXer:HARMonic:TYPE.....................................................................................872
[SENSe:]MIXer:HARMonic[:LOW]....................................................................................873
[SENSe:]MIXer:LOSS:HIGH............................................................................................873
[SENSe:]MIXer:LOSS:TABLe:HIGH.................................................................................873
[SENSe:]MIXer:LOSS:TABLe[:LOW]................................................................................873
[SENSe:]MIXer:LOSS[:LOW]...........................................................................................874
[SENSe:]MIXer:PORTs...................................................................................................874
[SENSe:]MIXer:RFOVerrange[:STATe]............................................................................ 874
[SENSe:]MIXer:FREQuency:HANDover <Frequency>
This command defines the frequency at which the mixer switches from one range to
the next (if two different ranges are selected). The handover frequency for each band
can be selected freely within the overlapping frequency range.
This command is only available if the external mixer is active (see [SENSe:]MIXer[:
STATe] on page 868).
Parameters:
<Frequency> numeric value
Example:
MIX ON
Activates the external mixer.
MIX:FREQ:HAND 78.0299GHz
Sets the handover frequency to 78.0299 GHz.
Manual operation: See "Handover Freq." on page 368
[SENSe:]MIXer:FREQuency:STARt?
This command queries the frequency at which the external mixer band starts.
Example:
MIX:FREQ:STAR?
Queries the start frequency of the band.
Usage: Query only
Manual operation: See "RF Start / RF Stop" on page 367