Remote Commands
R&S
®
FSW
934User Manual 1173.9411.02 ─ 19
11.8.3.3 Configuring and Performing a Marker Search
The following commands control the marker search.
CALCulate:MARKer:LOEXclude...................................................................................... 934
CALCulate<n>:MARKer:PEXCursion............................................................................... 934
CALCulate:MARKer:X:SLIMits[:STATe]............................................................................935
CALCulate:MARKer:X:SLIMits:LEFT................................................................................935
CALCulate:MARKer:X:SLIMits:RIGHT..............................................................................935
CALCulate:MARKer:X:SLIMits:ZOOM[:STATe]................................................................. 936
CALCulate:THReshold................................................................................................... 936
CALCulate:THReshold:STATe.........................................................................................936
CALCulate:MARKer:LOEXclude <State>
This command turns the suppression of the local oscillator during automatic marker
positioning on and off.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
CALC:MARK:LOEX ON
Manual operation: See "Exclude LO" on page 500
CALCulate<n>:MARKer:PEXCursion <Excursion>
This command defines the peak excursion.
The peak excursion sets the requirements for a peak to be detected during a peak
search.
The unit depends on the measurement.
Application/Result display Unit
Spectrum dB
Parameters:
<Excursion> The excursion is the distance to a trace maximum that must be
attained before a new maximum is recognized, or the distance to
a trace minimum that must be attained before a new minimum is
recognized
*RST: 6 dB in the Spectrum application and RF displays
Example:
CALC:MARK:PEXC 10dB
Defines peak excursion as 10 dB.
Manual operation: See "Peak Excursion" on page 500