Stereo Modulation (Option R&S SML-B5) R&S®SML / R&S®SMV03
1090.3123.12 E-6 4.32
MS
Des
cription: Sets or reads the music/speech flag.
The flag signals whether music or speech is being transmitted.
Command: MS=x
Query: MS?
Response: x
Value range: M|S
Example: Command:
STEReo:DIRect “MS=M“
The music/speech flag is set to "M". This signals that music is currently
transmitted.
Query:
STEReo:DIRect? “MS“
Response:
"M"
MPX-DEV
Description: Sets the MPX frequency deviation (max. deviation).
Command: MPX-DEV=xxxxx
Query: MPX-DEV?
Response: xxxxx
Value range: 00000 to 10000 (ASCII coded decimal numbers), corresponding to 0 Hz to 100 kHz
Example: Command:
STEReo:DIRect “MPX-DEV=00201“
Sets the MPX frequency deviation to 2.01 kHz.
Query:
STEReo:DIRect? “MPX-DEV“
Response:
"00201"
Note: A five-digit value must always be set. Leading zeros, if any, must also be specified.
PI
Description: Sets or reads the program identification (PI) code.
Command: PI=xxxx
Query: PI?
Response: xxxx
Value range: 0000 to FFFF (ASCII coded hexadecimal numbers)
Example: Command:
STEReo:DIRect “PI=1234“
The program identification code to be transmitted is set to "1234".
Query:
STEReo:DIRect? “PI“
Response:
"1234"
Note: A four-digit value must always be set. Leading zeros, if any, must also be specified,