IIT-2000/2010/5000/5010 Instruction Manual/English
90 05/08/2015 Version No. 001
View Sweep
Graph
Unlike normal manual tests, the special test mode also has an
option to view the resultant test as a sweep graph.
This option is only available for the
IIT-5000/5010.
Steps
1. When the test has finished, press
the SWEEP key to view the results
of the sweep in a graph.
W E ES P
Graph Test Items:
TEST
Primary
Secondary
ACW
Test voltage
measured current
DCW
Test voltage
measured current
IR
Test voltage
measured resistance
GB
Test current
measured resistance
DCW Example
D CW
Primary item
Secondary item
I ( m A ) :
V ( k V ) :
0 . 2 0 0
0 0 . 5 1
T ( S ) :
0 0 0 4 . 5
V
I
Test Time CursorSTA.t time

Table of Contents