3-12
Cleaning the Air Intake Grille
If dust and dirt are trapped in the air intake grille (for air-cooling the condenser) on the lower left side of the
instrument, they may cause the cooling performance to decline. Remove dust and dirt from the air intake
grille with a vacuum cleaner or something.
Replacement of the Brine Solution (auxiliary cooling fluid)
The brine solution must be replaced on a regular basis. If the brine solution continues being used even after
its replacement time, it will deteriorate and cause instrument malfunction. When the preset replacement
time comes close, the instrument displays the replacement notice automatically. Contact the nearest
authorized Sakura distributor or representative for replacement of the brine solution, when needed.