1-16
Specifications
Frozen Tissue Section Preparation Instrument
Cryostat Microtome
785 mm (W) x 781 mm (D) x 1185 mm (H) (650 mm (W) without handwheel handle)
500 mm (W) x 360 mm (D) x 280 mm (H) (360 mm (D) excludes the evaporator)
100 VAC 50/60 Hz 15 A, 115 VAC 50/60 Hz 13 A, 230 VAC 50/60 Hz 6.5 A
Specimen head (specimen holder): -10°C to -50°C (room temperature: 25°C)
Chamber: -10°C to -35°C (room temperature: 15°C)
-40°C or below (room temperature: 25°C)
Cryobar specimen capacity
-55°C or below (room temperature: 25°C)
Sectioning thickness setting
range
Blade angle adjustment range
Horizontal travel of blade holder
±20 mm from the specimen mounting base
Anti-Roll Plate, Anti-Roll Rake
Ozone disinfection cycle: Approx. 75 minutes (including 15 minutes of ozone decomposition)
Ozone concentration: Approx. 1000 ppm
Ozone decomposition: Manganese catalyst
Optional (HEPA filter capable of removing at least 99.99% of 0.3 µm or larger particles)
Standard equipment (pre-filter capacity: 164 cm
3
)
Motorized sectioning speed
6.2" VGA TFT color LCD display (640 x 240 dots)
Single-phase 100/115 VAC ± 10%, 50/60 Hz, 15 A or more, with grounding terminal
Single-phase 230 VAC ± 15%, 50/60 Hz, 7 A or more, with grounding terminal
Operating temperature: 10 to 35°C
Operating humidity: 30 to 85% RH (non-condensing)
Protection against electric shock
Class I Pollution Degree 2
* The equipment specifications are subject to change without notice.