2-28
Sectioning Automatically (DM only)
When the specimen has been cut to flat surface, perform sectioning.
Set the motorized sectioning mode to FINE or PAUSE.
Unlock the handwheel and press the pause key (
) and stop key () on the right panel together to perform
sectioning.
You can cut the specimen thinly (final shaving) to the preset section thickness.
Refer to p. 1-30 if you want to change the section thickness setting.
If the [MOTORIZED] Tab Screen is currently not displayed or the handwheel is not yet unlocked, an error
sound is output and motorized sectioning cannot be started.
Besides manual sectioning, you can use the Anti-Roll Rake or Anti-Roll Plate.
Refer to p. 2-12 for the details of manual sectioning.
Refer to p. 2-13 for information on how to use the Anti-Roll Rake.
Refer to p. 2-14 for information on how to use the Anti-Roll Plate.
How to Operate the Right Panel
Refer to p. 1-11 for information on how to operate the right panel.
How to Operate the Foot Switch
Refer to p. 1-10 for information on how to operate the foot switch.