2-17
After Processing
Cancelling Cleaning Operation
1. To cancel cleaning operation, you must enter the password.
2. Pressing the "Stop" button displays the cleaning operation
cancellation password confirmation screen.
Once the correct password is entered, the cleaning operation
cancellation confirmation screen appears.
If cleaning operation ends while this screen is displayed, the
screen closes automatically.
This password entry screen is for aborting the cleaning
operation, not for logging on the system.
3. Select "Abort," and defrost cycle operation will be cancelled if
the cleaning operation consists of the defrost cycle only, after
which the screen will close.
If the ozone cycle is also set as part of the cleaning operation,
the cleaning operation cancellation screen will appear and the
ozone cycle will be cancelled. However, the ozone purge cycle
cannot be cancelled.