Commissioning
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG 161
Subject to change without notice
5.7.3.4 Configuration of the Diagnostics Comparison limits
The Diagnostics Comparison limits can be activated to make the meter generate a warning
when the difference between the diagnostic values in the reference classes and those in
the current classes exceed the Diagnostics Comparison limit values.
These limits can be activated and configured in the User Warnings window:
Use MEPAFLOW600 CBM to connect to the meter (
pg. 125, 5.3).
Open the "User Warnings" window from the main system bar by clicking the "User"
button (
Figure 77).
Go to the "Diagnostic Comparison limits" tab (
Figure 77).
Use
Table 24 to plan the optimum configuration of the Diagnostics Comparison limits
to best suit your application.
Activate or deactivate the Diagnostic Comparison limits with the check box on the right.
Edit parameter values in the fields.
Click the "Write to meter" button.
All Diagnostics Comparison limit parameters can be configured in the User
Access Level "Authorized Operator" and without switching the meter to the
Configuration Mode.