Maintenance
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG 103
Subject to change without notice
7.2 Visual check of the MCS100E
7.3 Check sample gas flow
1 View the “bar graph” display (F2 key).
2 View the “flow” bar graph (if necessary, press the “right” cursor key).
3 Compare the flow to the nominal value (application-dependent).
7.4 Check/replace the filter pad in the fan of the analyzer
The fan is located on the underside of the MCS100E.
It is not necessary to switch the MCS100E off to replace the filter pad.
1 Pull the cover off.
2 Take filter pad out.
3 Clean the filter pad or insert a new filter pad.
4 Press the cover on.
What is to be checked Check
Ambient conditions Installation location, ambient temperature, air humidity,
corrosive atmosphere.
Piping Cracks, porous, brittle?
Enclosure and inside of device Soiled, corroded? Loose parts? Odor? Unusual noise?
Are service and fault messages shown on
the display?
----
Spare parts required Qty. Part No.
Filter pad for analyzer
1
1
As required
1 5309683
Risk of soiling.
Operate the MCS100E without filter pad only for a short period.