Commissioning
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG 39
Subject to change without notice
4.2 External keyboard
To ease the entry of rather large volumes of data a standard type external keyboard can be
attached to the keyboard connector (DIN-connector with 5 poles) at the door.
4.2.1 Country-specific assignment of keys
Depending on the type of external keyboard attached, it may be necessary to adjust the
country-specific assignment of keys.
1 Switch MCS100E off and then on again.
2 Abort automatic start after query <Start Program Y/N> (
p. 45, §5.3.3)
3 If you want to remain at DOS level for more than 4 minutes: Disable the watchdogs
(
p. 109, §7.8).
4 Start the “Norton Commander” at DOS level with: <c:\nc>
Note: Switch MCS100E off and on again before selecting the “Norton Commander”.
5 In the “Autoexec.bat” file (<F4>):
–Insertrem(blank)” before the line with the previous language and
delete “rem” before the line with the desired language.
Example: After switching from German (gr) to English (us):
rem C:\dos\keyb gr,,c:\dos\keyboard.sys
C:\dos\keyb\ us,,c:\dos\keyboard.sys
6 If required, enable watchdogs again (
p. 109, §7.8).
7 Exit the “Norton Commander”.
8 Start the measurement program with 1.bat (
p. 45, §5.3.2).
4.3 Language selection of the MCS program
The language of the MCS program is set:
In the MCSCONF configuration program “Directories and Files”
In the measurement sequence program in menu option Specialist: Utilities: Service:
Environment Configuration: Text file.
4.4 Display setup
Setting in the measurement program: (
p. 92, §5.7.15.7.1)
Setting via external keyboard:
CTRL ALT +
CTRL ALT –
The settings are not saved.
To enable the display of country-specific characters on the screen, country-
specific code pages must be loaded. This task should only be carried out by a
specialist who is well-acquainted with MS-DOS.