Specifications
136 MCS100E Operating Instructions 8009504/V YWA7/V3-1/2018-01 © SICK AG
Subject to change without notice
10.2.2 Operating data
The MCS100E equipment depends on the application. Please see the system
documentation enclosed with the MCS100E for the configuration
Measured value recording
Measuring principle Infrared photometer
Dual frequency and gas filter correlation method
Spectral range 1 .. 16 μm
Number of measuring components Max. 8 + O
2
Number of measurement ranges 2, automatic measurement range switch-over (adjustable)
Typical measuring components
NH
3
; HCl; H
2
O; SO
2
; CO; NO; NO
2
; N
2
O; CO
2
; NH
4
O
2
(ZrO
2
)
Interference compensation Max. 4 disturbance variables (also external disturbance variables)
Detection limit < 2% of respective measurement range
Zero drift < 1 % of the full-scale value / month
Sensitivity drift < ± 2 % of the full-scale value / month
Temperature effect < 2% of respective measurement range / 10 K
Setting time t
90
Application-dependent
Limit values 2 limit values per measured component
Device features
Enclosure dimensions (665 x 440 x 350) mm (HxWxD)
Enclosure material Steel plate, coated
Enclosure color RAL 7035 (light gray)
Weight approx. 70 kg
Window material CaF
2
(standard); Option: BaF
2
, IG2
Seal material Gas inlet: PTFE
Cell body: Viton Option: Kalrez
Cell window: Viton with FEP sheath; Option: Kalrez
Optical path length (cell) 3, 6, or 12 m
Sample gas volume (cell) approx. 2 l
Heating temperature (cell) typ. 100 °C or 185 °C, max. 200 °C
Sintered metal filter Pore size 10
μm
Internal sensitivity check Max. 3 (option)
Sample switching Max. 12 measuring points
Ambient conditions
Ambient temperature +5 … +35 °C
Storage temperature –20 … +60 °C
Relative humidity Max. 80% (non-condensing)
IP classification IP 52
Barometric correction 0.7 .. 1.2 bar ambient pressure (option)