Maintenance
104 MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG
Subject to change without notice
7.5 Span gas feeding
Feed a test gas with a defined test gas concentration (nominal value) and compare the
displayed value with the nominal value.
In case of deviation:
Check the gas path for leak tightness
Adjust the test gas value
If programmed: In the program (start with function button)
OR
–In the Specialist menu: System edit: Set Calibration data xxxx “Factor
OR
In the Specialist menu: Utilities: Service: Control Detector unit
When the drift is above the specification (see Technical Data): Please contact SICK
Customer Service.