EasyManuals Logo

SICK MCS100E User Manual

SICK MCS100E
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #110 background imageLoading...
Page #110 background image
Maintenance
110 MCS100E Operating Instructions 8009504/V YWA7/V3-1/2018-01 © SICK AG
Subject to change without notice
7.8.2 Watchdog for Communication
This watchdog controls the internal data communication.
Timeout: 4 minutes.
This watchdog cannot be disabled.
By removing a jumper it is possible to stop enabling the reset to the CPU.
Procedure
Put the MCS100E out of operation (
p. 100, §6.1) and disconnect from the power
supply.
Open the MCS100E.
Set the jumper of the watchdog:
Figure 22 Position and significance of jumper
•Close the MCS100E.
Put the MCS100E into operation again (
p. 45, §5.3).
WARNING: Hot surface - risk of burns
The surface of the cell insulation is hot.
Do not touch the surface of the cell insulation.
Jumper
.
.
Jumper to the left: Reset of Watchdog enabled
Jumper removed: Reset of Watchdog disabled
Before the new start of the measurement sequence program, set the jumper to
the “left” again.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK MCS100E and is the answer not in the manual?

SICK MCS100E Specifications

General IconGeneral
BrandSICK
ModelMCS100E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals