5.7.11 System edit (Name)
Specialist: System edit
This menu option contains the configuration for the system (= the measurement sequence)
name. The system to be configured can be:
• The actually active system. In this case any modifications to the settings have an
immediate effect on the configuration.
•A system that is not active at the moment (Selected in: Specialist: System, select for
editing). In this case, there is no effect on the actually active system. The system edited
can subsequently be activated via Specialist: System, activate for measurement.
5.7.11.1 Signal output
Specialist: System edit: Signal output
The following are defined here:
• Analog signal outputs
• Measurement ranges
Max. 32 outputs can be activated.
The menu option is identical to Main Menu: Configuration: Signal output
Settings
Channel Number of the analog output.
Mark () Analog output active ( )/not active.
Name Name of component derived from a list which is based on Specialist: System edit:
Results.
Display list by pressing <ENTER>
Unit Unit of component. Displayed automatically from Specialist: System edit: Results
Measurement
ranges
Switching of the measurement range is programmed here.
Measurement range 2 is the larger measurement range and relevant for the
plausibility check for zero setting- and for calibration, with the drifts etc.
The measurement ranges are automatically switched over (this is also true for the
graphic output). Switching over can be interrogated in the menu Specialist: System
edit: Zero/Calib/Purge programs/other programs
(Variable RHi).
Hysteresis is 5%.
If one component has been defined more than once, then the 1st definition is
relevant for the plausibility check.
Relays Number: Number of the relay
The relay “Number” is activated (deactivated) when the measurement range 2 (1) is
active (the relay indicates that the analog output switched to another output range).
0: No relay to be assigned
- Number: Inverted setting
Current 0..20 mA
4..20 mA