Maintenance
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG 105
Subject to change without notice
7.6 Programming the Measurement Sequence (Overview)
1 Insert a new result
Number of result: --> System documentation.
Menu: Specialist: System edit: Results
2 Enter filter wheel positions
Menu: Specialist: System edit: Definition of Components
3 Define calibration data
Menu: Specialist: System edit: Calibration data
4 Shift and insert linearization tables
Menu: Specialist: System edit: Linearization tables
5 Shift and insert IF tables
Menu: Specialist: System edit: Interference table
6 Define graphics
State the components and measurement ranges you wish to be displayed on the
screen.
Note: Do not simply append the new component at “the rear”. Program the components
in the same sequence as defined in “Results”.
Menu: Specialist: System edit: Graphic output
7 Define analog outputs
Leave the extension of analog outputs to SICK Customer Service.
Note: Do not simply append the new component at “the rear”. Program the components
in the same sequence as defined in “Results”.
Menu: Specialist: System edit: Signal output
8 Define data storage and send to PC
Specify the desired components
Menu: Specialist: System edit: Data storage and send to PC
9 If you want to calibrate the new component manually: Enter the component in the
Softkey menus.
Menu: Specialist: System edit: Softkey menu definition
10 Supplement programs
Menu: Specialist: System edit: Zero/Calib/Purge programs/other programs.
11 If a new span gas valve was fitted: Supplement Operator's Notebook
Menu: Specialist: System edit: Operator's Notebook
Assign a suitable name to the digital output used for the span gas valve (e.g.: do11: Y8
Cal Gas NO)
12 Perform text and relay assignments
Menu: Specialist: System edit: Messages and corresp. relays
NOTICE:
Wrong entries have a disruptive effect or cause nonfunctioning of the
measurement program.
Leave changes of the measurement sequence to trained personnel.