SV 977D User Manual - Appendixes
• Linear measurement with the Vel10 and MFVel filters:
from 0.145 mms
-1
to 5.13 ms
-1
(the sinusoidal signal RMS)
from 0.205 mms
-1
to 7.25 ms
-1
(PEAK)
Table C.2.5. Linear operating ranges with the Vel10 filter
Linear operating ranges (RMS)
With the SV 80 (10 mV/ms
-2
) accelerometer sensitivity @ 79.58 Hz
Measured at the ref. environmental conditions, at 10 Hz sine signal
• Linear measurement with the Dil1 filter
from 248 µm to 8.81 m (the sinusoidal signal RMS)
from 351 µm to 12.45 m (PEAK)
Table C.2.6. Linear operating ranges with the Dil1 filter
Linear operating ranges (RMS)
With the SV 80 (10 mV/ms
-2
) accelerometer sensitivity @ 79.58 Hz
Measured at the ref. environmental conditions, at 1 Hz sine signal
• Linear measurement with the Dil3 filter
from 24.8 µm to 880 mm (the sinusoidal signal RMS)
from 35.1 mm
to 1.245 m (PEAK)
Table C.2.7. Linear operating ranges with the Dil3 filter
Linear operating ranges (RMS)
With the SV 80 (10 mV/ms
-2
) accelerometer sensitivity @ 79.58 Hz
Measured at the ref. environmental conditions, at 3 Hz sine signal
• Linear measurement with the Dil10 filter
from 2.3 µm to 81.6 mm (the sinusoidal signal RMS)
from 3.25 µm
to 115.4 mm (PEAK)
Table C.2.8. Linear operating ranges with the Dil10 filter
Linear operating ranges (RMS)
With the SV 80 (10 mV/ms
-2
) accelerometer sensitivity @ 79.58 Hz
Measured at the ref. environmental conditions, at 10 Hz sine signal