7 Maintaining the product
190
7.9. Carrying out a software and firmware system
update
- To make the most of the Saveris system performance,
the system should be updated regularly.
- Software and firmware updates should always be carried
out together. For updates to versions up to 4.2 always
begin with the software update. For updates to versions
4.3 and above always begin with the firmware update.
- All software and firmware updates are available on the
Testo website in the download centre or from our
software hotline (softwarehotline@testo.de).
- For further help carrying out the updates, contact our
software hotline softwarehotline@testo.de
Pos: 208 /TD/Produ kt instand halten/te sto Saveris/08 b Software- Update durchf ühren PROF @ 12\mod_1338384491723 _79.docx @ 126050 @ 2544 @ 1
7.9.1. Carrying out a software update
- The Saveris measuring and configuration data is
retained during this process, as it is stored on the SQL
Server®.
- If MAPI or SMTP Mail is installed, the settings should be
documented in the registry under
HKeylokalmachine\software\testo\comsoft\tdasmail and
the email components under the software should be
uninstalled before the Saveris server uninstallation and
reinstalled after the Saveris server installation.
- If the storage path for your automated reports have
been updated in the registry, these settings should be
updated again in the registry.
- After a Saveris CFR software update, the security
settings in the software must be reconfigured. For this,
please remove the old configuration settings, close the
software, open it again and reassign the security
settings in the 3 Testo groups.
Uninstall the older version of the Saveris client (Professional or
CFR version) without uninstalling the database and then install the
current server.
7.9.1.1. Uninstalling software
1. Change to Control panel - Programs and features in
Windows®.
2. In Programs and features, click on Testo Saveris
Professional Server.