• Res:Meas:T
Returns the currently measured average temperature value as 1.23456e-3. This value
depends on
which temperature channels are marked for average calculation.
- Query: Res:Meas:T?
[TAverage]
• Res:Meas:Sequence
Returns the current measuring sequence for the started resistance measurement. The
parameter
winding describes whether no (-) or only primary (1) or only secondary (2) or both (D)
windings will be
measured. Mode defines which supply mode will be used. In case that no measurement
will be
performed for a sequence step, mode will be set to “Nop”. And phase describes between
which
terminals the resistance will be measured for primary and secondary winding.
- Query: Res:Meas:Sequence?
[Winding],[Mode],[PhasePri],[PhaseSec];
[Winding],[Mode],[PhasePri],[PhaseSec];
[Winding],[Mode],[PhasePri],[PhaseSec];
[Winding],[Mode],[PhasePri],[PhaseSec];
[Winding],[Mode],[PhasePri],[PhaseSec];
[Winding],[Mode],[PhasePri],[PhaseSec];
[Winding1] = - ¦ 1 ¦ 2 ¦ D
[Mode] = Nop ¦ 32A/50V ¦ 16A/100V ¦ 16A/50V ¦ YdnMode ¦
Classic2R ¦ UnknownOperation
[Phase] = A-N ¦ B-N ¦ C-N ¦ N-A ¦ N-B ¦ N-C ¦
A-B ¦ B-C ¦ C-A ¦ B-A ¦ C-B ¦ A-C
• Res:Meas:State
Returns the current resistance measurement state. SeqNo indicates which measurement
out of the
sequence described above is currently performed. An RxStable is set to 1 if the measured
resistance
value is stable according to the Res:Set:Data:Stability criterion.
- Query: Res:Meas:State?
[SeqNo],[R1Stable],[R2Stable],[R3Stable]