196
Remote Control
the values of HV and LV, respectively the values of HV and TV, depending on which
windings are
selected. In classic mode R1, R2 and R3 can be obtained. The same holds for the
measured voltages.
- Query: Res:Meas:R?
[R1],[R2],[R3],[R4]
Res:Meas:R? 2
[R2]
• Res:Meas:I
Returns the currently measured current values as 1.23456e-3. In automatic mode I1 and
I2 are the
values of HV and LV, respectively the values of HV and TV, depending on which windings
are
selected. In classic mode I1 is the measuring current for all resistors.
- Query: Res:Meas:I?
[I1],[I2]
Res:Meas:I? 1
[I1]