MAINTENANCE AND CHECKING
REPLACING THE FUSE
1 Make sure the power is OFF and the power cable is unplugged.
2 Press the fuse holder with a screwdriver and turn it counterclockwise.
The fuse holder can be removed.
3 Replace the fuse with the proper fuse provided with the instrument.
4 Press the fuse holder lightly with a screwdriver and turn it clockwise.
The fuse holder is now reset.
WARNING
To avoid electric shock when replacing the fuse, be sure to
unplug the instrument before removing the fuse cover. Do not
use ungrounded outlets. Do not plug in the instrument without
the fuse cover.
Replacing the fuse