41
BASIC OPERATIONS
6
When the initial setting is Auto mode, and when the MEASUREMENT screen is displayed, a
message "Get the eye aligned and press the MEASUREMENT switch" is displayed on the
screen: the waiting status starts.
In this case, alignment operation is not done even when the setting of alignment mode is
"Auto."
7 Input the patient ID as needed. Tap the to display the patient data
input screen. For how to operate the patient data input screen, see step 2 and 3 in page
135, "SETTING THE PATIENT ID".
PREPARATION BEFORE MEASUREMENT
1 Make sure the MEASUREMENT screen is on.
The message disappears by pressing the or by touch-
ing the screen, following the message.
When the initial setting of alignment mode is "Auto," alignment operation starts.
Patient ID can be input at any time until the measurement data output.
CAUTION
Never insert your fingers under the chinrest.
* Inform the patient of this, too.
Careless insertion of fingers may cause injury by pinching.
CAUTION
To avoid electric shock, do not touch the external connection
terminal and the patient at the same time.
MEMO
The measurement in auto mode may not be possible due to the eyelid
and the eyelashes over to the pupil.
If this occurs, the operator should tell the patient to open their eyes wide
enough, or lift the eyelid to allow for measurement.
MEMO
Measurement in auto mode may not be possible due to frequent blinks
or existing abnormalities in the corneal surface caused corneal disease
etc.
In this case, use manual mode.
MEMO
When operating the instrument, be careful not to touch the patient's face
or nose. If touched, clean the instrument following "
CLEANING THE
INSTRUMENT" on page 180.
MEMO
The automated alignment function might not work appropriately when
the patient is doing the shining make-up around her eyelid.
In this case, please change into the manual mode of operation.
MEASUREMENT switch
Patient ID button